SRN1C-02

Single-chanel analyzer

It is used in isotopic radiometry, in radioimmunoanalysis of biological samples and in radiopharmacy.


Composition of the SRN1C-02 radiometric system
Well counter:
– Single-channel radiometer with USB port to PC, model SRN1C-02.
– Scintillation detector holder with NaI(Tl) well crystal.
– USB connection cable.
– Power cord.
Application Software.
User manuals.
CD with software, driver, manuals and utilities.

Technical specifications
Features of the measuring probe:Well detector
Thallium-activated sodium iodide (NaITl) scintillation crystal
50 x 50 mm well scintillating crystal
Well dimensions: 30 x 15 mm
High voltage source for the detectors:Output Vmax = 1200 V (adjustable)
Software controlled in 256 interval
Output Imax = 1 mA
Energy range:10 keV- 200 keV
Maximum measurement time:60 000 s
Minimum measurement time:1 s
Sensitivity:between 3 and 16 %
Resolution:>10 % for Am 241
Detection efficiency:> 60 %
Maximum activity:10 µCi (370 kBq)
Mains:110 V– 240V CA
650 mA (max.)
50-60 Hz
Dimensions:Radiometer: 235 x 170 x 88 mm
Detector holder: Ø = 100 mm, l= 270 mm
Net weight:9.6 kg